Philips BV Pulsera C Arm Specifications
X-Ray System
Specialty Packages Available
Extended processing.
Vascular functions.
Extended vascular functions.
Cardiac Pulsed functions.
Generator
80 kHz High frequency.
7.5 kW full-wave.
Up to 120 kVp.
Up to 60 mA for radiographic film exposures.
Continuous fluoro boost mode up to 20 mA.
Pulsed fluoro mode.
Pulsed fluorography mode up to 60 mA.
Full power from outlet:
50 Hz or 60 Hz.
100V-20A to 240V-10A.
X-Ray Tube
Rotating anode X-ray tube.
Focal spots: 0.3 mm and 0.6 mm.
Anode heat capacity: 300,000 HU.
Anode cooling rate: 70,000 HU/min.
Housing heat capacity: 1,900,000 HU.
Anode diameter: 2.8 in.
Anode angle: 10°.
Collimator
Iris collimation with dual opposing lead shutters allowing elliptical, round and rectangular collimation.
Minimum beam size < 5 X 5 cm2.
Iris and shutters are continuously variable.
All functions remotely controlled from the C-Arm control panel.
Fluoroscopy Mode
Focal spot: 0.3 mm and 0.6 mm.
kVp range: 40 – 120 kVp.
mA range: 0.2 – 20 mA continuous High Definition fluoro boost mode.
Auto and manual modes.
Continuous or Pulsed operation.
APF varies mA, kVp and camera gain.
User specific APF tables.
Pulsed Fluoroscopy Mode
Half Dose Fluoroscopy.
Quarter Dose Fluoroscopy.
Focal spot: 0.6 mm.
kVp range: 40 – 120 kVp.
mA range: 0.4 – 12 mA.
Pulse rate: 6.25 or 12.5 pulses per second.
Pulse width: 11 to 37 milliseconds.
Computer controlled APF, mA, kVp and camera gain.
Pulsed Fluorography Mode
Focal spot: 0.6 mm.
kVp range: 40 – 110 kVp.
mA range: Up to 60 mA.
Pulse rate: 3 to 30 pulses per second. (optional)
Pulse width: 8.0 / 9.5 / 11.1 milliseconds.
Computer controlled camera iris, mA, kVp, & camera gain.
Digital Exposure Mode
Focal spot: 0.6 mm.
kVp range: 40 – 110 kVp.
mA range: 1.0 – 75 mA.
Radiographic Mode
Focal spot: 0.6 mm.
kVp range: 40 – 120 kVp.
Fixed 60 mA.
0.6 mm - mAs range: 3.2 – 125 mAs.
Video Imaging System
Image Intensifier
9 in Image Intensifier
Tri-mode 9 in/7 in/5 in image intensifier
9 in.: 1.6 lp/mm
7 in.: 2.1 lp/mm
5 in.: 2.9 lp/mm
DQE: 65% (typical)
12 in Image Intensifier (optional)
Tri-mode 12 in/9 in/7 in image intensifier
12 in.: 1.2 lp/mm
9 in.: 1.4 lp/mm
7 in.: 2.0 lp/mm
DQE: 65% (typical)
Video Camera
High resolution CCD camera.
525 line/frame.
Full frame capture.
360° motorized rotation.
Left-right image reversal.
Top-bottom image reversal.
Negative mode.
Computer controlled features: (optional)
Gain.
Blanking.
Camera iris.
Video Monitors
Dual 17 in high resolution, anti-glare, anti-static.
Ambient room light sensor.
Multi-LED X-ray “ON” indicator.
Hardcopy Options
Thermal printer. (optional)
DICOM. (optional)
Image Capture Device. (optional)
Laser aiming device. (optional)
C-Arm Dimensions
Height: 71 / 91 in. (1805 / 2305 mm)
Width: 35 in. (880 mm)
Length: 76 in. (1930 mm)
Weight: 695 lbs 315 kg.
Workstation Dimensions
Height: 69 in (1760 mm)
Width: 32 in (800 mm)
Depth: 37 in (925 mm)
0777201555